TGIC temperature gradient cross chromatographic analyzer

TGIC is the abbreviation of temperature gradient interaction high temperature chromatography, which uses the adsorption and desorption of polyolefin chains and TGIC columns and the combined effect of crystallization properties to achieve the separation and characterization of different components of polyolefin at a certain solvent flow rate and temperature change.

TGIC temperature gradient cross chromatographic analyzer

Classification:

Characterization of analytical instruments

Key words:

Product optimization

Detailed introduction:

TGIC temperature gradient cross chromatography product introduction

TGIC is the abbreviation of temperature gradient interaction high temperature chromatography, which uses the adsorption and desorption of polyolefin chains and TGIC columns and the combined effect of crystallization properties to achieve the separation and characterization of different components of polyolefin at a certain solvent flow rate and temperature change. In particular, some non-crystalline parts that cannot be analyzed by crystallization classification technology can be analyzed by TGIC. The analysis of special polyolefins such as elastomers and block polyolefins is the main analysis field of TGIC.

1. TGIC function

1) Equipped with a fully automatic sampler to analyze 42 samples, and each sample can choose a different analysis method;

2) Information on the chemical composition of the sample and its distribution and information on the comonomer content, including the soluble fraction, can be obtained. The chemical composition and distribution of the soluble part and the CCD characterization of special polyolefins such as block polymers are not possible with traditional crystallization techniques TREF and CEF;

3) The CH3/1000C of the sample can be directly obtained by selecting IR4 component probe or IR5;

4) The viscosity detector and laser light scattering detector are selected for component-molecular weight interaction analysis;

5) Select IR5 MCT high sensitivity detector to improve the sensitivity of concentration signal;

6) With remote control and interactive diagnosis function.

2. TGIC Features

1) Automatic analysis, automatic completion of solvent addition, sample dissolution and online filtration and analysis process, simple hardware;

2) The analysis process does not require manual contact with solvents and meets HSE requirements;

3) Using a dedicated software program, TGIC and CEF can use the same instrument for analysis;

4) Good repeatability;

5) The instrument software design is very friendly and easy to operate;

6) The instrument is easy to use and maintain.

3. Application Example

Application Example 1

The sample of a scientific research institute is made by blending polypropylene and POE. Through conventional chemical composition analysis instrument TREF analysis, the blend has two components at 12.4 ℃ and 115.3 ℃, but most of them are soluble. Traditional TREF analysis cannot give more effective information. The results of sample analysis by TGIC are shown in fig. 2:TGIC can separate and characterize the components with little difference in crystallization performance, obtain the chemical composition and distribution information of the sample, and obtain the branching degree information under different components, thus improving the conventional chemical component characterization means and solving the analysis problem of POE materials.

Fig.1 Sample TREF analysis results

Fig.2 TGIC analysis results of samples

Application Example 2

For the analysis of block polyolefin ipp and PE, the traditional TREF and CEF techniques cannot clearly characterize their chemical composition and distribution, and cannot draw differences (see Figure 3). After the temperature gradient interaction high temperature chromatography TGIC analysis, you can get very valuable results, as can be seen from Figure 4, the more block polyethylene content, the stronger the adsorption capacity of the column, the more the fraction will be washed out at a higher temperature, so as to understand the number of blocks and whether there is a block.

Fig.3 Results of traditional TREF and CEF characterization techniques

Fig.4 Schematic diagram of TGIC characterization results

Application Example 3

The analysis of ethylene octene copolymer samples is a typical case of TGIC application. The traditional CEF and TREF analysis can not get the results, but the results obtained by TGIC can be seen that the octene content decreases with the increase of elution temperature. At 40 ℃, the octene content of the retained component is as high as 32%.

Fig. 5 TGIC analysis of ethylene octene copolymer samples