Equipment
SKY1003-IVA automatic kinematic viscosity tester
SKY1003-IV automatic kinematic viscosity tester, suitable for the determination of crude oil, paraffin, light and heavy fuel oil, lubricating oil, additives, oil and other transparent or opaque kinematic viscosity of petroleum products. The test accuracy meets or exceeds the requirements of ASTM D445.
Classification:
Petrochemical Products
Key words:
Product optimization
Detailed introduction:
Product Introduction
SKY1003-IV automatic kinematic viscosity tester, suitable for the determination of crude oil, paraffin, light and heavy fuel oil, lubricating oil, additives, oil and other transparent or opaque kinematic viscosity of petroleum products.
The test accuracy meets or exceeds the requirements of ASTM D445.
Applicable standards: ASTM D445, ASTM D446, GB/T265, GB/T30515, ISO3104, ISO3105, IP71
Product Features
- The instrument adopts iSKvator technology, automatic sampling, constant temperature, viscosity testing, cleaning and drying viscometer, without random operation
- Advanced NTC thermistor measurement technology, the test is not interfered by sample color and ambient brightness, and can test transparent, dark and opaque
- Ming petroleum products to the same precision
- Improved Ubelohde capillary viscometer with independent intellectual property rights, each viscometer has a test range of 100 times
- Kinematic viscosity measurement range: 0.5 ~ 10000cSt
- Automatic preheating of waxy oil samples or heavy oil
- Dual solvent cleaning system, user-programmable cleaning program
- Use dry air to dry the viscosity tube to ensure the accuracy of the test
- Corrosion resistant design, can accept most corrosive cleaning solvents
- Only about 10ml of test oil sample
- The whole cleaning process requires at least 10-15ml of cleaning solution.
- The whole process of sample testing and cleaning only takes about 10min
- The sample injection tray can be moved forward as a whole, and the protective cover of the transparent oil suction nozzle is installed in a card type, which is convenient for users to clean up the residue of the sample absorption regularly.
- Software with temperature calibration function, viscosity tube calibration function and viscosity index automatic calculation function
- 12-bit automatic sample injection tray, after the software is registered and sorted, it can be unattended for continuous automatic testing, or it can be used for priority testing.
- According to the characteristics of the experimental oil sample, the test and cleaning templates in the software can be modified to meet the individual needs of the test.
- Real-time record of experimental data, easy to browse the experimental results and query the historical data
- Can be connected to LIMS system to transmit data
Technical parameters
|
Technical parameters |
|
|
Test Method |
GB/T 265,GB/T 30515,ASTM D445, ISO 3104, IP 71,ASTM D446, ISO 3105. |
|
Temperature control range |
20 ℃ ~ 100 ℃ (special cold source shall be selected for 20 ℃) |
|
Temperature control accuracy |
± 0.01 ℃ |
|
Timing accuracy |
0.02s |
|
automatic injection |
12-position automatic sample tray |
|
Sample preheating |
Room temperature ~ 95 ℃ |
|
Constant temperature bath capacity |
2.5L |
|
Constant temperature bath medium |
Dimethicone |
|
Viscosity test range |
0.5~10000 mm2/s (by selecting different viscosity tubes) |
|
Test accuracy |
At or better than the level specified in ASTM D445 |
|
capillary viscometer |
modified ub viscometer |
|
Security |
Cleaning fluid lack of tips, bath temperature overheating protection, high and low liquid level protection |
|
Communication Interface |
Host: Two RS232 ports |
|
Data Output |
LIMS or Printer |
|
Overall power |
≤ 1000W |
|
Heating power |
600W |
|
Instrument use environment |
Temperature: 5 ℃ ~ 35 ℃ |
|
Instrument power supply |
AC 220V±10 50/60Hz |
|
Overall dimension |
400mm wide x 510mm deep x 710mm high |
|
net weight of instrument |
60KG (including constant temperature medium) |
Related recommendations