Equipment
GPC-QC03
Simplified and fully automated GPC/SEC instruments for quality control laboratories in polyolefin production plants
Classification:
Online Mall
Key words:
Product optimization
Detailed introduction:
1. Fully automatic sample preparation and analysis system
1) Fully automatic sample preparation: manual only need to weigh the solid sample, and then put the sample into the sample bottle of the autosampler. With in-line filter and backwash function, no external filtration is required.
2) Automatic analysis process: After the sample is placed, there is no need to directly contact the solvent to protect the health of the operator. After the analysis is completed, the relevant pipelines are automatically cleaned to wait for the analysis of the next sample, which is fully in line with HSE standards.
3) New design to reduce degradation: the autosampler is divided into normal temperature area and high temperature area. Generally, the sample is located in the normal temperature area and the temperature is controlled by the program. The sample bottle is transferred to the high temperature area in turn according to the software program to ensure the same heating time of each sample. The sample is oscillated and dissolved to reduce shear degradation. Equipping nitrogen purge accessories can deoxidize nitrogen and minimize high temperature oxidation degradation.
4) On-line flow marker addition system: This system is selected and can automatically add flow markers. After its peak appears, the analysis program has been completed.
2. Infrared detector (IR4 or IR5MCT)
1) A special selective infrared detector, IR4 or IR5MCT detector only responds to C- H infrared absorption, with stable baseline, high selectivity and good repeatability.
2) high sensitivity: it can meet the MMD analysis of very dilute polyolefin solution, at a concentration of 0.125 mg/ml, the signal to noise ratio can reach 60.
3) While measuring the molecular weight and its distribution, the value of the number of kilo-carbon methyl groups or the information of short chain branches can also be directly measured, especially for the analysis of HDPE pipe material with low number of kilo-carbon methyl groups (the detection limit of IR5MCT detector components is less than 1); Combined with a viscosity detector, the information of long chain branches can also be obtained.
Related recommendations
Message consultation